File name 34405A Firmware Revision History Agilent_34405A_Firmware_Revision_History c20140214 [4].pdf Date Description Revision History 6/2/2006 34405A firmware update to 1. Solve the firmware download 1.19-3.05 (Changed from V110- issue for a particular batch of V303) new ATMEL IC. 2. Error 631 on InGuard is resolved. 6/28/2006 34405A firmware update to 1. Error 635 will pops up if user 1.20-3.05 download only the MB firmware. Reason is the OPC code was not sent after the reset. Solution ---- sent the OPC code. 2. Once in a while the CAL checksum is cleared while performing calibration (either manual or remote) Reason is the one public variable? |